Manufacturer Specifications:
Features:           
  • Portable, compact, flexible and convenient to use
  • Can measure, display and record standard 12-lead ECG waveforms.
  • 6 channel with measurement report
  • 5.6-inch single-color LCD: high light and high definition
  • can display electrocardiogram wave and rhythm of heart and catch record opportunity
  • Built-in rechargeable Ni-MH battery. Non-memory rechargeable battery to ensure continuous use of the device 
  • Chinese/English language.
  • Measuring storage and review functions. Automatically saves records when user finishes printing every time. Each record has a maximum time of 50 seconds.
  • Against high frequency electrosurgical generator equipment and defibrillation-proof
  • Record means: automatic channel connecting means, manual channel
  • Connecting means; 6-lead, 3+1-lead, 3-lead electrocardiogram, long time rhythm channel connecting record and periodical record.
  • Real time record: record waves for 10s according to channel connecting
  • 50/63/110/210mm recorder, can record 3-lead and 6-lead electrocardiogram.
  • Web recording paper is more convenient
  • Can be used to the physiological measure for adult, child and neonate.
Screen:

Size

 
Catercorner 5.6 inches
Type Single Color
LCD
Resolution Ration
320*240

 

Control:

Criterion

 
4 direction keys (F1-F4), Power supply switch key, Storage key, Print Start/Stop key.

 

 
 
EKG:

Usable standard leads

 
Ⅰ,Ⅱ,Ⅲ, V1, V2, V3, V4, V5, V6, Avr, Avl and Avf
Lead fail
Identify the failed lead.
Alarm
No alarm
Input Impedance:
≥25MΩ
Voltage Range:
±300Mv
Common mode rejection ratio
≥90Db
Frequency Response:
0.05Hz-150Hz
Standard Precision:
10mm /1Mv±2%

 

 Paper Recorder:

Method:

 
Dot matrix
Level Resolution Ration:
40 d/mm, speed 25mm/s
Uprightness Resolution Ration:
8 d/mm
Wave Channels:
3/6
Paper Width:
110mm(4.33inches)
Paper Length:
15m(50inches)
Formfeed Speed:
5, 10, 12.5, 25 and 50mm/s(±2%)

 

Environment Instructions:

Power Supply Requirement:

 
AC:120V~240V 50Hz/60Hz
Battery:
NI-MH, 12V, 1.8AH(model:205305)
Fuse:
2.0A, F2L 250V
Design(Routine):
Continuous operation, non-ingress of liquids Device.
Battery life:
When in ECG monitoring, the normal working time are 3 hours for fully charged battery.( Battery aging will reduce the using time)
Battery charged for more than 90%charging time:
6 ~ 12 hours
Operation Conditions:
Surrounding Temperatures:
Relevant humidity:
Atmospheric Pressure:
 
10 ~ 40
25
~95%(No Condensation)
700 ~ 1060hPa
Restore Conditions:
Surrounding Temperature:
Relevant humidity:
Atmospheric Pressure:
 
-20℃~55℃
≤93%
500hPa ~1060hPa