Manufacturer Specifications:
Features:
- Folded 12.1-inch TFT LCD high-brightness screen with a wide viewing angle and large numerics
- 12 elements / 1 MHz water proof pulse Doppler transducer for accurate detection
- Event marker for easy documentation of events and kick counts
- Fetal Movement Profile (FMP) detects and automatically records gross fetal body movements
- On-screen scrolling for viewing stored fetal traces
- Twin ultrasound monitoring capability
- Configurable audible and flashing fetal heart rate alarms
- Twin signal overlap alarm function
- Built-in high speed thermal printer, printing width can be set to 112mm,150mm and 210mm
- 24-hours memory for fetal traces
- Supporting TCP/IP or wireless connection to the central monitoring station
- Internal Li-ion battery power or AC power from external power supply
|
Physical Specifications |
||||
|
Dimension and Weight
|
Dimension
|
35cm (W)*30cm (D)*11cm (H)
|
||
|
Weight
|
standard configuration: 4 kg
|
|||
|
Operation Environment
|
Power requirements
|
AC100-240V , 50/60Hz
|
||
|
Power
|
100VA
|
|||
|
Temperature
|
5-40℃
|
|||
|
Humidity
|
30-93% non-condensing
|
|||
|
Performance Specifications
|
||||
|
Performance Specifications
|
Display
|
12.1inch (800X600 pixel) color TFT –LCD
|
||
|
Indicator
|
Visual on-display and integrated, alarm light, audible alarm
|
|||
|
Battery
|
Internal rechargeable Li-ion battery power
|
|||
|
Alarm
|
Visual on-display and integrated, alarm light, audible alarm
|
|||
|
Memory
|
24-hours memory for fetal traces
|
|||
|
Printer
|
America standard/international standard
Built-in thermal printer (110mm, 150mm, 210mm for option)
1/2/3cm/min real-time printing speed, fast print speed (stored traces)
|
|||
|
Paper
|
Z-fold, thermosensitive
|
|||
|
Record Message
|
ID, name, age, date, time, print speed etc
|
|||
|
CMS
|
Support connecting to central monitoring station through LAN (TCP/IP) or wireless connection
|
|||
|
Technical Specifications
|
||||
|
Ultrasound (FHR)
|
Measurement range
|
US
|
50 to 240 bpm
|
|
|
Resolution
|
Display
|
1bpm
|
||
|
Printer
|
1bpm
|
|||
|
jitter@200 bpm
|
≤3bpm
|
|||
|
Display update rate
|
0.25second
|
|||
|
Beat to bea change(max.) for ultrasound
|
28bpm
|
|||
|
US frequency
|
1MHz±5%Hz
|
|||
|
TOCO
|
Measurement method
|
Strain gauge sensor element
|
||
|
Sensitivity
|
1 unit =2.5g
|
|||
|
Resolution
|
1 unit
|
|||
|
Measurement range
|
500 units
|
|||
|
Baseline setting
|
0, 5,10, 15, 20 can be selective
|
|||
|
Update rate
|
0.25 second
|
|||
|
Auto offset correction
|
3 second after connecting the transducer
|
|||
|
Auto zero adjust
|
TOCO value is set to zero negative following a negative
Measure value for 5 seconds
|
|||
|
Auto Fetal Movement (AFM)
|
Technique
|
Pulsed doppler ultrasound
|
||
|
Range
|
0-100%
|
|||
|
Resolution
|
1%
|
|||
|
Automatic fetal movement actograph
|
||||
|
Manual fetal movement mark
|
||||
|
NIBP
|
Operation mode
|
Manual/automatic/ STAT
|
||
|
Measurement unit
|
mmHg / kPa selectable
|
|||
|
Test interval
|
0-480 minutes
|
|||
|
Accuracy
|
Max.Std.deviation:8 mmHg(1.1kPa)
|
|||
|
Max. mean Error: ±5mmHg(±0.7kPa)
|
||||
|
Measurement range
|
Systolic
|
30 to 270 mmHg(4 to 36 kPa)
|
||
|
Diastolic
|
10 to 245 mmHg(1.5 to 32kPa)
|
|||
|
Mean
|
20 to 255 mmHg(2.5 to 34kPa)
|
|||
|
Pulse rate range
|
40~240bpm
|
|||
|
Over-pressure protection
|
Yes
|
|||
|
Maternal ECG (MECG)
|
Heart rate range
|
30~300bpm
|
||
|
Resolution
|
1bpm
|
|||
|
Accuracy
|
±1bpm/ ±1%
|
|||
|
Leads
|
I, II, III, avL, avR, avF, V
|
|||
|
Sweep speed
|
12.5mm/s, 25mm/s, 50mm/s
|
|||
|
Gain selection
|
x 0.25, x0.5 ,x1,1, x2
|
|||
|
Operation mode
|
diagnostic, monitor, surgery
|
|||
|
Maternal SpO2 (MSpO2)
|
Measurement range
|
0%-100%
|
||
|
Resolution
|
1%
|
|||
|
Accuracy
|
±2%
|
|||
|
Pulse rate range accuracy
|
±3bpm
|
|||
|
Temperature (TEMP)
|
Technique
|
Thermistor probe CYSI 400 series compatible
|
||
|
Measurement range
|
0℃-50℃
|
|||
|
Accuracy
|
±0.1℃
|
|||
Previous:
Next:




View larger image
Bookmark this product
Add to My Favorites